• 4 Principle
  • ISO 27608:2010(E) 2 © ISO 2010 – All rights reserved 5 Apparatus 5.1 Automatic colorimeter
  • State standard of the republic of uzbekistan




    Download 287.21 Kb.
    Pdf ko'rish
    bet4/15
    Sana16.06.2023
    Hajmi287.21 Kb.
    #73664
    1   2   3   4   5   6   7   8   9   ...   15
    Bog'liq
    27608-разблокирован
    DALINA, Abdulla Qahhor. Paqir qo\'mondonlari
    3 Terms and definitions 
    For the purposes of this document, the following terms and definitions apply. 
    3.1 
    Lovibond
    ®
     colour value 
    value obtained by analysis of light transmitted through a known path length of liquid fat or oil and read from 
    the display of a suitable instrument 
    NOTE Lovibond
    ®
    colour value is displayed in terms of Lovibond
    ®
    colour units. 
    4 Principle 
    The colour of the light transmitted through a known path length of liquid fat or oil is measured by the spectral 
    response of the light arriving at a detector. The results of these calculations are expressed in terms of 
    Lovibond
    ®
    colour units. 
    1) Lovibond is the trade name of a colour scale developed and supplied by The Tintometer Ltd 
    (
    http://www.tintometer.com
    ). This information is given for the convenience of users of this document and does not 
    constitute an endorsement by ISO of the product named. Equivalent products may be used if they can be shown to lead to 
    the same results. 
    O‘z DSt ISO 27608:2017 (ISO 27608:2010, IDT)


    ISO 27608:2010(E) 

    © ISO 2010 – All rights reserved
    5 Apparatus 
    5.1 Automatic 
    colorimeter. An instrument capable of measuring transmitted colour over the range 400 nm 
    to 700 nm, reporting the result in terms of Lovibond
    ®
    colour units, being equivalent to those obtained using the 
    manual equipment according to ISO 15305
    [5]
    . It should also be capable of giving results equivalent to 
    ISO 15305
    [5]
    when optical glass cells ranging from 1,58 mm to 133 mm (1/16 inch to 5 
    1/4
    inch) optical path 
    length are used. 
    The following instruments have been found to be suitable: 
    a) LICO 200, LICO 300, LICO 400, and LICO 500
    2)

    b) PFX 880/L, PFX 950, and PFX 995
    3)


    Download 287.21 Kb.